4) PH 7 calibration Connect the PH ELECTRODE with the " BNC socket " ( 3-10, Fig. 1 ) and immerse the electrode in the PH7buffer solut

Pagina 3 - 1. FEATURES

CAL4.00 * The texts " CAL " will flash for around 5 seconds.After that, the meter calibrate itself automatically.The upper display will show

Pagina 4 - 2-1 General Specifications

4-5 OthersAbove calibration procedures effect only when reading valuewithin 1 PH of the calibration point.However if the reading value beyonds *1 PH

Pagina 5

5. PH TEMPERATURE COMPENSATIONEnable the meter to gain high accuracy measuring resultsfrom different kinds of solution, the temperature compensation c

Pagina 6

a. Adjust the temperature compensation value * Press the " TEMP. C Button " ( 3-7, Fig. 1 ) first, The upper display will show the measured

Pagina 7 - 3. FRONT PANEL DESCRIPTION

@ Use the " Up Button " ( 3-5, Fig. 1 ) and the " DownButton " ( 3-8, Fig. 1 ) to adjust the PH 4 defaultcalibration value.@ The

Pagina 8

The upper display will show the values of " 10.00 " andthe lower display will show the " PH 10 defaultcalibration value ".PH10.001

Pagina 9 - 4. PH CALIBRATION PROCEDURE

5-2 Automatic temperature compensation1) Plug the " Optional ATC Temp. Probe, YK-200ACT "into the " Optional Probe Input Socket "

Pagina 10

4) * If the operating is under the " Manual temperaturecompensation ", then please refer to the above 5-1calibration procedures. * If the o

Pagina 11 - 4-4 Singal Point Calibration

6-4 Data Hold Press the " Hold Button " ( 3-4, Fig. 1 ) will hold the measured value & the LCD will indicate a " HOLD " symbol

Pagina 12

TABLE OF CONTENTS1. FEATURES... 12. SPECIFICATIONS...

Pagina 13

6-6 Following are the block diagrams for quick measuring procedures CalibrationSet the Single point calibrationPower manual PH 7(CAL.)ON temp.valu

Pagina 14

Power managementAUTO POWER OFF MANUAL POWER OFF(Not activated during orMemory Record Selection)7. DISABLE AUTO POWER OFFThe instrument has " Auto

Pagina 15

Meter PC(3.5 mm jack plug) (9W 'D" Connector)Center Pin... Pin 4 Pin 22.2 K Ground/shield...

Pagina 16

9. BATTERY REPLACEMENT1) When the left corner of LCD display show " ", itis necessary to replace the battery. However, in-spec.me

Pagina 17 - 6. MEASURING PROCEDURE

CONDUCTIVITY* YK-200PCD plug into thePROBE YK-2001PH to become aprofessional Conductivity Meter.Model :YK-200PCDMeasurementRangeConductivity 2 mSCondu

Pagina 18 - 6-3 Temp. Measurement

ATC * YK-200PATC plug into thePROBE YK-2001PH to be a ATC (Automatic TemperatureCompensation ) Probe for PHModel : function.YK-200PATCMeasurementRange

Pagina 19

1. FEATURES* Master meter is a professional PH/mV METER.PH range : 0 to 14 PH x 0.01 PH.mV range : -1999 mV to 1999 mV.* DPHC2001 can be plugged with

Pagina 20 - measuring procedures

* 蚓 or 蚌 can be converted by push button on the front panel.* PH calibration is easily to be done by push button on thefront panel.* Using the durable

Pagina 21 - 8. RS232 PC SERIAL INTERFACE

Intelligent * The instrument can be plugged with thefunctions optional Conductivity and TDS probe(YK-200PCT), Conductivity probe(YK-200PCD), Dissolve

Pagina 22 - RS232 FORMAT : 9600, N, 8, 1

Weight 250 g/0.55 LB ( battery included ).Size ( meter ) 195 x 68 x 30 mm ( 7.6 x 2.6 x 1.2 inch ).Standard Instruction manual...1 PC.Accesso

Pagina 23 - 9. BATTERY REPLACEMENT

3. FRONT PANEL DESCRIPTIONFig. 15

Pagina 24

Fig. 2 Fig. 33-1 Display3-2 Power Button3-3 REC. Button3-4 Hold Button3-5 C, F Button/A Button/up Button3-6 PH/mV Button/B Button/left Button3-7TEMP.

Pagina 25 - Measuremen

4. PH CALIBRATION PROCEDURE 4-1 Calibration Consideration The most ideal PH ELECTRODE generates 0 mV at PH7.00 ( 177.4 mV at PH 4 ) and YK-2001PH ha